Datastick pH Meter

For use with the AquaSensors system, The pH DataStick connects directly to a computer or PLC(Programmable Logic Controller) for direct integration with industrial control systems. This allows data to be displayed and analised, and for the instrument to be customised and calibrated without the need for intermediate controller box. The pre-calibrated sensor heads can be replaced or exchanged with any other type of sensor without any system downtime, saving time and money.

The datastick is designed to be extremely strong, non-flowing, resistant to fouling and with exceptional chemical resistance. The internal electronics of the sensor are completely encapsulated and O-ring sealed, giving complete protection from humidity and moisture. An integral temperature sensor automatically compensates measurements for changes in process temperature, and a titanium ground electrode eliminates ground loop currents in the measuring electrode.

DataStick Measurement System
pH DataStick Specifications
Performance
Range 0 to 14 pH
Resolution 0.01 pH
Accuracy 0.1% of reading
Step Response Time 90% in 30 seconds
Operational Environment - PEEK Sensor Head
Temperature Range ?5 ?C to 95 ?C
Maximum Pressure 100 psig @ 95 ?Ce
Maximum Flow Rate 10 ft/second
Operational Environment - CPVC Sensor Head
Temperature Range ?5 ?C to 95 ?C
Maximum Pressure 85 psig @ 75 ?C
Maximum Flow Rate 10 ft/second
Power Requirements
Voltage Range 10 to 30 VDC
Maximum Power 200 mW
Typical Power 120 mW
Construction
Process Electrode ?G?, ?HF?, low temperature, pure water, or high temperature glass
Ground Rod Titanium, 316 stainless steel or Hastelloy C
O-rings : Viton? (other materials available)
Sensor Head Material PEEK or CPVC
DataStick Material 316 stainless steel, PEEK, or CPVC
Weight 1.2 lbs (PEEK or CPVC)
2.6 lbs (316 stainless steel)
Units of Measure
Measurement Units pH, mV
Temperature Units ?C, ?F
Calibration
Automatic Buffer 1 and 2 point
Sample 1 and 2 point
Temperature 1 point
Temperature Compensation Options
Linear % per ?C
Built-in tables Ammonia or Morpholine
Other Configuration Options
Sensor Filter 0 to 100 seconds
Temperature Filter 0 to 100 seconds
Auto Calibration Buffer Standards : (4, 7, 10) and DIN 19267
Approvals and Ratings
Immunity & Emissions CE Certified 89/336/EEC:CISPER 11, EN61000 (-4-2,-4-3,-4-4,-4-6, 4-8)
Safety cULus Listed; 367G E303570
Hazardous Locations Haz Loc Class 1, Division 2, Groups A, B, C, D. Max Ambient 50?C
Download Product Brochure Download PDF

Select Options