| TS3000 Specification |
| Measurement technique |
90? Infra-red (ISO7027) |
| Std ranges available (factory set) |
100NTU, 400NTU and 1,000NTU
|
| Sensor Outputs |
Internal Data Logger ? serial data via Smartcom software
Optional adaptor provides SDI12 serial output (3 wire) |
| Resolution |
0.1NTU 0.2NTU 0.3NTU |
| Linearity |
+/- 1% +/- 1% +/- 3% |
| Calibration standard |
APS AEPA polymer solutions 0,100NTU,400NTU,1000NTU |
| Cable type |
Polyurethane sheathed cable, OD 8mm, with 3mm vent tube, moulded entry |
| Cable lengths |
10, 20, 30, 50, 100, 150m (32, 65, 100, 165, 325, 490 ft) |
| Power supply |
9 to 30Vdc (at sensor), or on-board battery pack (option) |
| Power ESD protection |
2000 volts |
| Sensor warm up time |
10 seconds |
| Battery capacity |
9 x Lithium AA (3.6Volt) ? Total capacity 5.2Ah @ 10.8v |
| Typical field life (battery pack) |
Over 6 months remote operation @30 min data logging. |
| Internal data logger |
Non-volatile, battery backed RAM with real time clock |
| Memory size |
2Mb capacity, with user selectable wrap function |
| Data storage |
250 000 readings. (Typically 5 minute data for >12 months) |
| Logging frequency |
User selectable from 1 second up to once per day |
| Averaging alarms, |
User selectable via supplied Smartcomm software |
| Depth rating (water column) |
100m |
| Storage temperature |
-5?C - +60?C |
| Weight |
500g plus cable weight (665g per 10m length) |
| Dimensions (L x OD) |
418mm x 47mm (16.46? x 1.85?) |
| Wetted materials |
UPVC, acetal,316 passivated stainless steel, polyurethane, viton |
| Wiper arrangement |
Disposable foam pad on PVC arm. Field replaceable |
| Wiper control |
User selectable via software |
| Wipe time |
6 seconds (1 revolution) |