HOBO 4-Channel Thermocouple Data Logger

Product Options
UX120ON014M

Overview

The HOBO UX120 Thermocouple Data Logger is a four-channel LCD data logger for measuring and recording temperature in a broad range of monitoring applications. The data logger makes it easy and convenient to record temperatures over a broad range (-260 to 1820? C) and can accept up to four J, K, T, E, R, S, B, or N type probes.

In addition to accepting four thermocouple probes, the data logger features an internal temperature sensor for logging ambient temperatures, further extending the range of application possibilities.

Onset also offers a single-channel thermocouple data logger, the HOBO UX100-014M.

Manufactured by: Onset
Model: HOBO 4-Channel Thermocouple Data Logger

Services offered for this product

  • Sales
  • Rental
  • Installation
  • Calibration
  • Maintenance

Applications

Temperature

Specification

J Type Thermocouple
Range -210° to 760°C (-346° to 1,400°F)
Accuracy ±0.6°C (±1.08°F) ± thermocouple probe accuracy
Resolution 0.03°C (0.06°F)

K Type Thermocouple
Range -260° to 1,370°C (-436° to 2,498°F)
Accuracy ±0.7°C (±1.26°F) ± thermocouple probe accuracy
Resolution 0.04°C (0.07°F)

T Type Thermocouple
Range -260° to 400°C (-436° to 752°F)
Accuracy ±0.6°C (±1.08°F) ± thermocouple probe accuracy
Resolution 0.02°C (0.03°F)

E Type Thermocouple
Range -260° to 950°C (-436° to 1,742°F)
Accuracy ±0.6°C (±1.08°F) ± thermocouple probe accuracy
Resolution 0.03°C at (0.05°F)

R Type Thermocouple
Range -50° to 1,550°C (-58° to 2,822°F)
Accuracy ±2.2°C (±3.96°F) ± thermocouple probe accuracy
Resolution 0.08°C (0.15°F)

S Type Thermocouple
Range -50° to 1,720°C (-58° to 3,128°F)
Accuracy ±2.2°C (±3.96°F) ± thermocouple probe accuracy
Resolution 0.08°C (0.15°F)

B Type Thermocouple
Range 550° to 1,820°C (1,022° to 3,308°F)
Accuracy ±2.5°C (±4.5°F) ± thermocouple probe accuracy
Resolution 0.1°C (0.18F)

N Type Thermocouple
Range -260° to 1,300°C (-436° to 2,372°F)
Accuracy ±1.0°C (±1.8°F) ± thermocouple probe accuracy
Resolution 0.06°C (0.11°F)

Internal 10K Thermistor (Temperature)
Range Logging: -20° to 70°C (-4° to 158°F)
Accuracy ±0.21°C from 0° to 50°C (±0.38°F from 32° to 122°F), see Plot A in manual
Resolution 0.024°C at 25°C (0.04°F at 77°F); see Plot A in manual
Drift <0.1°C (0.18°F) per year
Logger
Logger Operating Range
Logging -20° to 70°C (-4° to 158°F); 0 to 95% RH (non-condensing); logger accuracy from 0° to 50°C (32° to 122°F)
Launch/readout 0° to 50°C (32° to 122°F) per USB specification
 
Logging rate 1 second to 18 hours, 12 minutes, 15 seconds
Logging modes Normal, Burst or Statistics
Memory modes Wrap when Full or Stop when Full
Start modes Immediately, Push Button, Date & Time or Next Interval
Stop modes When Memory Full, Push Button, or Date & Time
Restart mode Push Button
Time accuracy ±1 minute per month at 25°C (77°F), see Plot B in manual
Battery life 1 year, typical with logging rate of 1 minute and sampling interval of 15 seconds or greater
Battery type Two AAA 1.5V alkaline batteries, user replaceable
Memory 4 MB (1.6 million measurements, maximum)
Download type USB 2.0 interface
Full memory download time Approximately 1.5 minutes
LCD LCD is visible from: 0° to 50°C (32° to 122°F); the LCD may react slowly or go blank in temperatures outside this range
Size 10.8 x 5.41 x 2.54 cm (4.25 x 2.13 x 1 in.)
Weight 107.5 g (3.79 oz)
Environmental rating IP50
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU)

Downloads