The Aquistar TempHion multiparameter water quality instrument with integrated power and datalogger is capable of measuring pH, specific ions (chloride, bromide, and nitrate), redox and temperature. One thermistor based temperature element comes with each unit, plus up to four pH, ISE, or redox elements.
The TempHion is powered internally by two AA batteries or with an external auxiliary power supply for data intensive applications.
The TempHion water quality sensors can be programmed using a laptop or desktop via it's RS485/RS232 adapter and easy to use Aqua4Plus software.
After programming, the sensor will measure and collect data at a variety of time intervals.
The internal processor allows easy calibration, using the Aqua4Plus calibration utilities. Once calibration is complete, this data is stored in the non-volatile memory within the Smart Sensor. This calibration information is applied to recorded data, resulting in highly accurate readings at a wide range of temperatures.
Manufactured by: INW
Model: TempHion Multi-Parameter Water Quality Sensor
Saltwater intrusion tracking
Wastewater treatment discharge
pH monitoring
Tidal influence studies
Single or multi well tracer tests
General | |
Length | 17.625" (44.8 cm) |
Diameter | 0.75" (1.9 cm) |
Weight | 0.85 lb. (0.4 kg) |
Body Material | Delrin® & 316 stainless steel or titanium |
Wire Seal Material | Viton® and PTFE/FEP |
Submersible Cable | Polyurethane, polyethylene, or FEP |
Protection Rating | IP68, NEMA 6P |
Terminating Connector | Available |
Communication | RS485 Modbus® RTU SDI-12 (ver.1.3) |
Direct Modbus Read Output | 32-bit IEEE floating point |
Millivolt Channels | No. Available Channels: 2mv, 1 temp Range: ± 1200mV Accuracy: 0.1% value Resolution: 0.1mV |
Reference | Electrode: Ag/AgCl solid-state electrode Junction: Patented capillary liquid junction Electrolyte: TempHionTM reference solution |
Operating Temp. Range | -0° C to 55° C |
Storage Temp. Range1 | -20° C to 80° C |
Maximum Depth | 700 ft (210 m) | 300 PSI |
Logging | |
Memory | 2MB - 200,000 records |
Log Types | Variable, user-defined, logarithmic, profiled |
Programmable Baud Rate (optional) | 9600, 19200, 38400 |
Logging Rate | 2x/sec maximum |
Software | Complimentary Aqua4Plus or Aqua4Push |
Networking | 32 available addresses per junction w/ batching capabilities (up to 255) |
File Formats | .xls / .csv / .a4d |
Power | |
Internal Battery | 2x1.5V AA alkaline2 |
Auxiliary Power | 12VDC - Nominal 6-15VDC - Range |
Exp. Alkaline Battery Life | 18 months at 15m polling interval3 |
Temperature | |
Element Type | 30K thermistor |
Element Material | Epoxy bead/external housing |
Accuracy | ± 0.2° C |
Resolution | 0.1° C |
Range | -5° C to 60° C |
Units | Celsius, Fahrenheit, Kelvin |
pH/ORP | |
Sensor Type/Material | pH: Glass combination electrode ORP: Platinum ring |
Ranges | pH: 0-14 pH units / -538mV to 260mV ORP: ± 1200mV |
Units | pH, mV, Eh |
Typical Accuracy | pH: ± 0.2 pH units ORP: 0.1 mVH |
Resolution | pH: 0.01 pH units ORP: 0.01 mVH units |
Compensated Range | 0° C to 40° C |
Calibration | pH: One or two point calibration w/ pH buffers (7 & 4 or 10) ORP: EH 1 pt. calibration |
Referance Solution | Potassium Nitrate - (KNO3) |
1 Storage without batteries 2 Lithium available upon request 3 May vary due to environmental factors |