Moisture Target Series 6 Moisture Analyser (MTS6)

Product Options

Overview

The Moisture Image Series 6 (MTS6) moisture analyser measures moisture in compressed air and gases. It’s a single channel analyser which is designed for use in permanent applications. The MTS6 is compatible with M Series moisture probes or VeriDri moisture transmitters.

Manufactured by: Panametrics
Model: Moisture Target Series 6 Moisture Analyser (MTS6)

Services offered for this product

  • Sales
  • Rental
  • Installation
  • Calibration
  • Maintenance
  • IoT Cloud Compatible

Applications

Regenerative desiccant air or gas dryers
Plastic drying applications prior to injection blow molding
Medical air dryers
Glove box monitoring

Specification

Electronics
Intrinsic safety External safety barrier for moisture input (optional)
European compliance Complies with EMC directive 2004/108/EC and 2006/95/ EC low voltage directive (Installation category II, pollution degree II)
Input Moisture signal from Baker Hughes Sensing thin-film aluminum oxide moisture sensor on an M Series probe or VeriDri transmitter
Analog output Single internal isolated recorder output for dew point, internally optically isolated, 10-bit (0.1%) resolution
Standard switch-selectable outputs
  • 0 to 2 V, 10k W minimum load resistance
  • 0 to 20 mA, 400 W maximum series resistance
  • 4 to 20 mA, 400 W maximum series resistance
  • User-programmable within the range of the instrument and the corresponding sensor or transmitter
Alarm relays
  • One fail-safe fault relay
  • Two standard Form C relays SPDT, rated for 3 A at 250 VAC/30 VDC
  • Set to any level within the range of the instrument; programmable from the front panel
Alarm set point repeatability ±0.2°F (±0.1°C) dew point
Datalogger 32 GB capacity with MicroSD card, 2 GB card included
Display 128 x 64 matrix LCD
Display functions Dew point temperature in °F, °C or sensor signals for diagnostics
Power requirements Universal power 100-240 VAC @ 50-60 Hz, or 24 VDC nominal
Temperature
  • Operating: –4° to 140°F (–20° to 60°C)
  • Storage: –40° to 158°F (–40° to 70°C)
Warm-up time Meets specified accuracy within three minutes
Configurations
  • Panel (front panel meets Type 4 and IP66)
  • Board version, without enclosure
Moisture Measurement
Sensor type Thin-film aluminum oxide
Moisture probe compatibility Compatible with all Baker Hughes M Series aluminum oxide moisture probes and VeriDri transmitters
Dew/frost point temperature
  • Overall range capability: 140°F to –166°F (60°C to –110°C)
    • Standard 68°F to –112°F (20°C to –80°C) with data to -166°F (–110°C)
    • Extended 140°F to –85°F (60°C to –80°C) with data to 166°F (–110°C)
  • VeriDri ranges include:
    • –110°C to 20°C
    • –110°C to –50°C
    • –90°C to 10°C
    • –80°C to –30°C
    • –80°C to 20°C
    • –30°C to 20°C
    • –130°F to 70°F
    • –130°F to –40°F
    • –40°F to 70°F
    • –50°F to 50°F
    • –100°F to 0°F
Calibrated accuracy at 77°F (25°C)
  • ±3.6°F (±2°C) from –85°F to 50°F (–65°C to 10°C)
  • ±5.4°F (±3°C) from –112°F to –86°F (–80°C to –66°C)
Repeatability
  • ±0.9°F (±0.5°C) from –85°F to 50°F (–65°C to 10°C) dew/frost point
  • ±1.8°F (±1.0°C) from –112°F to –86°F (–80°C to –66°C) dew/frost point

Downloads