Orion VERSA STAR Conductivity Module

Orion VERSA STAR Conductivity Module

  • Warranty This product is covered by a three year meter warranty
  • Large Memory Capable of holding up to 2000 data points
  • Module Offers Many Reading Options
Product Options

Overview

RS Hydro offer the Orion VERSA STAR Conductivity Module which offers a choice of practical salinity or natural sea water curves for salinity readings and selectable reading reference temperatures of 5, 10, 15, 20 or 25°C with the option of linear, non-linear, nLFu or EP curve for conductivity readings.

There are many features that make the VERSA STAR Conductivity Module the ideal choice and they are;

  • Easy to understand screen prompts with soft keys that update for easy selection
  • The multi-language interface supports English, Spanish, French, Italian, German and Chinese
  • A large memory capacity holds up to 2000 data points all of which time and date stamps
  • Two USB and RS232 ports for easy transfer of data and updating of meter software
  • Data analysis software for transferring stored results or displaying readings in real time
  • Direct control for two Orion Star stirrer probes for mixing samples without a stir plate
  • IP54 rated housing protects from splashes
  • This product comes complete with a three year replacement warranty

Manufactured by: Thermo Scientific
Model: Orion VERSA STAR Conductivity Module

Services offered for this product

  • Sales
  • Rental
  • Installation
  • Calibration
  • Maintenance

Specification

Accuracy (Conductivity) 0.5% of reading ±1 digit
Accuracy (Resistivity) 0.5% of reading ±1 digit
Accuracy (Salinity) ±0.1ppt
Accuracy (TDS) 0.5% of reading ±1 digit
Calibration Points Up to 6
Cell Constant 0.001 to 199.9
Compatible Probes 2-cell or 4-cell conductivity with built in temperature
Data Points 2000
Description Orion VERSA STAR Conductivity/TDS/Resistivity/Salinity/Temperature Module
Inputs 8-pin Mini-DIN (conductivity with temperature probe)
Log Edit Type Delete individual points, delete group of points or delete all
Log Function Type Manual, ready (includes AUTO-READ), timed
Range (Conductivity) 0.001 µS/cm to 3000 mS/cm
Range (Resistivity) 2 ohm to 100 meg-ohm
Range (Salinity) 0.01 to 80.0 ppt NaCl equivalent; 0.01 to 42 ppt practical salinity
Range (TDS) 0 to 200ppt
Relative Humidity Range 5 to 85 %, non-condensing
Resolution (Conductivity) 0.001 µS minimum, 4 significant digits minimum
Resolution (Resistivity) 2ohms-cm
Resolution (Salinity) 0.01ppt
Resolution (TDS) 4 significant digits
Sensor Types 2-cell or 4-cell conductivity probe - purchase separate
Accuracy (Temperature) ±0.1°C
Temperature Compensation Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off
Range (Temperature) -5 to 105°C
Resolution (Temperature) 0.1°C
Voltage 90 to 260VAC
Warranty 3 year meter replacement
Salinity Measurement Modes Practical salinity or natural sea water (NaCl equivalent)
TDS Factor Range Linear 0.01 to 10.00
Certifications/Compliance IP-54
Item Description Conductivity/TDS/Resistivity/Salinity/Temperature Module
Temperature (Metric) Reference 5, 10, 15, 20, 25°C

FAQ's

Total dissolved solids (TDS) is a measure of the combined inorganic and organic substances dissolved in water. It is directly related to the electrical conductivity of water and as such can be used as an indication of change.
TDS is only relevant in freshwater systems as salinity will dramatically increase the conductivity and therefore apparent TDS.
TSS are particles that are larger than 2 microns found in the water column. Anything smaller than this is called a dissolved solid.