CyberScan DO 600

  • DO Measurement Range Dissolved oxygen measurement range from 0.00 to 90.00 mg/L or ppm
  • Built-In Datalogger Memory to store up to 500 data sets stamped with time and date
  • Dustproof and Waterproof IP67 rated housing makes it ideal for harsh conditions
Product Options
ECDOWP60041K

Overview

The CyberScan DO 600 has mega storage space capable of saving up to 500 data sets time and date stamped all of which are automatically recorded by the auto-logging function. All information is delivered on the comprehensive one-glance-tells-all screen complete with intelligent screen prompts.

The CyberScan DO 600 measures oxygen concentration up to 90.00mg/L and saturation up to 600.0%, this is one of the widest measuement ranges in the DO handheld market. Salinity, temperature and pressure compensation results are accurate even under varying conditions with the benefit of a built in barometer with auto pressure correction to always deliver accurate results. The simple step by step calibration guide enables the user to follow simple prompts for convenient 2 point calibration. There is also a handy cal-due alarm on real time clock which ensures periodic calibration updates.

The transfer of data is a simple process using IrDA wireless communication for direct meter to PC data transfer.

Housed in a rugged IP67 rated waterproof casing this handheld is ideal to be used in harsh environments in the field.

This product is covered by a 3 year meter warranty.

Manufactured by: Eutech Instruments
Model: CyberScan DO 600

Services offered for this product

  • Sales
  • Rental
  • Installation
  • Calibration
  • Maintenance

Applications

Aquaculture
Industrial
Environmental
Educational

Specification

Dissolved Oxygen Range Range: 0.00 to 90.00 mg/L or ppm
Resolution: 0.01 mg/L or ppm
Accuracy: ±0.2 mg/L
% Saturation Range: 0.0 to 600.0%
Resolution: 0.1%
Accuracy: ±2 %
Temperature Range: 0.0 to 50.0 °C (32 to 122 °F)
Resolution: 0.1 °C (0.1 °F)
Accuracy: ±0.3 °C (±0.5 °F)
Compensation: ATC / MTC (0.0 to 50 °C)
Salinity Correction Range: 0.0 to 50.0 ppt
Resolution: 0.1 ppt
Method: Automatic correction after manual input
Pressure Correction Range: 450 to 825mmHg (59.9 to 109.9 kPa)
Resolution: 1 mmHg (0.1 kPa)
Method: Automatic correction with inbuilt sensor
Calibration 2 Point (0%, 100%)
Operating Range 0.0 to 50.0 °C (32 to 122 °F)
Probe Type Galvanic
Date/Time Stamping Yes
Memory 500 data sets (GLP-compliant format)
Special Functions Auto-Hold; Auto Power-Off after 2 - 30 minutes (user-preset); Stability Indicator; Self-diagnostics messages
Power 4 x 1.5 'AA' batteries (>400 hours) or 9V DC adapter, 500 mA

Downloads

Accessories

Related products