Datastick ORP REDOX Meter

The DataStick System with ORP Sensor provides differential ORP measurement with no field calibration required. It can be connected directly to a PLC(Programmable Logic Controller), and any computer can be used to display data and to customize and calibrate the sensor. The unit is offered in a variety of materials and has replaceable quad junction salt bridges. Pre-calibrated sensor heads can be replaced or exchanged with any other type of sensor without any system downtime, saving time and money.

The sensor is constructed to be exceptionally chemically resistant and mechanically strong, and has non-flowing and anti-fouling characteristics. It also has a built-in pre-amplifier and universal engineering units conversion. It also has universal signal conditioning electronics and interactive communications with a host computer or display interface via a range of connectivity options, including Modbus? RTU, DeviceNet, Profibus, USB, CANopen or Ethernet.

A titanium ground electrode is included to eliminate ground loop currents in the measuring electrode and an integral temperature sensor measures temperature independently.

DataStick Measurement System
ORP Measurement System Specifications
Performance
Range ?2100 mV to +2100 mV
Resolution 0.1 mV
Accuracy 0.1% of reading
Step Response Time 90% in 30 seconds
Operational Environment - PEEK Sensor Head
Temperature Range ?5 ?C to 95 ?C
Maximum Pressure 100 psig @ 95 ?Ce
Maximum Flow Rate 10 ft/second
Operational Environment - CPVC Sensor Head
Temperature Range ?5 ?C to 75 ?C
Maximum Pressure 85 psig @ 75 ?C
Maximum Flow Rate 10 ft/second
Power Requirements
Voltage Range 10 to 30 VDC
Maximum Power 200 mW
Typical Power 120 mW
Construction
Process Electrode Platinum or gold
Ground Rod Titanium (standard), 316 stainless steel or Hastelloy C
O-rings Viton? (other materials available)
Sensor Head Material PEEK or CPVC
DataStick Material 316 stainless steel, PEEK, or CPVC
Weight 1.2 lbs (PEEK or CPVC)
2.6 lbs (316 stainless steel)
Units of Measure
Measurement Units mV
Temperature Units ?C, ?F
Calibration
Sample 1 point
Temperature 1 point
Temperature Compensation Options
Temperature is provided as an independent measurement
Other Configuration Options
Sensor Filter 0 to 100 seconds
Temperature Filter 0 to 100 seconds
Approvals and Ratings
Immunity & Emissions CE Certified 89/336/EEC:CISPER 11, EN61000 (-4-2,-4-3,-4-4,-4-6, 4-8)
Safety cULus Listed; 367G E303570
Hazardous Locations Haz Loc Class 1, Division 2, Groups A, B, C, D. Max Ambient 80?C
Download Product Brochure Download PDF