The AquiStar CT2X Conductivity Datalogger is a submersible conductivity / temperature sensor with built in datalogger. The CT2X conductivity datalogger is ideal for use in rugged field applications, as it incorporates a 4-pole epoxy/graphite electrode cell. This technology helps reduce fringe field interference errors, decreases inaccuracy caused by polarization effects and lowers contact resistance problems.
The conductivity data logger can store over 349,000 records of conductivity, temperature, and time data and has a non volatile memory which ensures data is not lost in the case of a power failure.
The CT2X is built from 316 Stainless Steel (or Titanium), Viton and PTFE/FEP with a diameter of just 0.75in or 19mm.
The conductivity dataloggers have low power consumption and needs just two AA batteries to power it which can be replaced by the user. Alternatively an external auxiliary power supply can be used for data intensive or SDI-12 applications.
The CT2X electrical conductivity datalogger has a linear and non-linear temperature compensation feature, however specific linear temperature coefficients can be entered to match the material being sampled. A non-linear natural water mode is available for water without linear temperature coefficients.
The CT2X conductivity logger comes with Aqua4Plus software, allowing the user extensive control, flexible programming and real time monitoring, as well as versatile display options. Data can be easily exported to spreadsheets and databases.
The CT2X logger is available with or without cable. Choices of polyurethane, polythene and PTFE/FEP are available.
Other features:
Manufactured by: INW
Model: CT2X Water Conductivity/Temperature Datalogger
Salinity ingress
Borehole monitoring
Conductivity studies
General | |
Length | 12.5” (31.8 cm) 16.0” (40.6 cm) w/ pressure |
Diameter | 0.75” (1.9 cm) |
Weight | 1.0lb. (0.5 kg) |
Body Material | Delrin® & 316 stainless or titanium |
Wire Seal Materials | Viton® and PTFE/FEP |
Submersible Cable | Polyurethane, polyethylene, or FEP |
Cable Weight | 4lbs./100 ft (1.8 kg/30 m) |
Protection Rating | IP68, NEMA 6P |
Desiccant | 1-3mm indicating silica gel (high or standard capacity) |
Terminating Connector | Available |
Communication | RS485 Modbus® RTU SDI-12 (ver.1.3) |
Operating Temp. Range3 | -15°C to 55° C |
Storage Temp. Range1 | -40°C to 80° C |
Logging | |
Memory | 4MB - 349,000 records |
Log Types | Variable, user-defined, logarithmic, profiled |
Programmable Baud Rate | 9600, 19200, 38400 |
Logging Rate | 4x/sec maximum |
Software | Complimentary Aqua4Plus or Aqua4Push |
Networking | 32 available addresses per junction w/ batching capabilities (up to 255) |
File Formats | .xls / .csv / .a4d |
Power | |
Internal Battery | 2x1.5V AA Alkaline2 |
Auxiliary Power | 12VDC - Nominal 6-15VDC - Range |
Exp. Alkaline Battery Life | 12 months at 15m polling interval7 |
Temperature | |
Element Type | 30K ohm thermistor |
Element Material | Epoxy bead/external housing |
Accuracy | ± 0.2°C |
Resolution | 0.1°C |
Range | -5°C to 60° C |
Units | Celsius, Fahrenheit, Kelvin |
Conductivity | |
Probe Material | Epoxy/Graphite |
Electrode | 4-pole |
Static Accuracy | ± 0.5% of measured value |
Resolution | 32 bit |
Ranges | Conductivity6: 0-100,000 or 0-200,000 or 0-300,000 µS/cm TDS: 4.9-49,000 or 4.9-98,000 or 4.9-147,000 mg/L Salinity: 2-42 PSU |
Units | µS/cm, mS/cm, mg/L, PSU |
Resolution | 0.1 µS/cm / 0.001 mS/cm / 0.1 mg/L (TDS) / 0.001 PSU |
Warm-Up Time | 200 msec |
Thermal Compensation | None, linear, or nLFn |
1Storage without batteries 2 Lithium available upon request 3Requires freeze protection kit if using pressure option in water below freezing 4Burst reduced at PSI>300 5Higher pressure ratings available upon request 6Accuracy reduced at levels <10 µS/cm 7May vary due to environmental factors |