Overview

The 6712 Portable Wastewater Sampler features a vacuum formed ABS plastic shell which is able to withstand the harshest of conditions. The ISCO 6712 portable sampler has a trim 20 inch diameter design allowing it to be easily installed and removed from a manhole. Transporting this portable wastewater sampler is safe and convenient due to the large comfortable handles. The NEMA 4X, 6 (IP67) enclosure is submersible, watertight, and dust-tight. The Plug-in 700 Series Modules and the SDI-12 interface make it easy to add flow and parameter monitoring to your 6712-based system.
The advanced 6712 Controller allows the user to choose from a selection of programming modes to ensure the most suitable program for the application.

Programming is made fast and simple.

Standard Programming mode allows the user is walked through the sampling sequence step by step, allowing the user to the correct parameters specific to the application. The Extended Programming Mode on this portable wastewater sampler lets the user enter more complex programs including ?smart sampling? event notification, triggered on any combination of up to 16 inputs.

MCERTS

The Isco 6712 Portable Waste Water Sampler is now MCERTS certified meeting the Environment Agency's performance requirements. View downloads for certificate.

Standard Features

  • Choice of 11 different glass and plastic bottle configurations ranging from 24 x 1liter to 1 x 5.5 gallon.
  • NEMA 4X, 6 (IP67) controller enclosure
  • Rugged ABS plastic shell
  • Foam-insulated base holds up to 30 pounds (13.5 kg) of ice to preserve samples even in extreme conditions.
  • Sample delivery at the EPA-recommended velocity of 2 ft/sec., even at head heights of 26 feet
  • SDI-12 interface provides plug and play connection with multi-parameter water-quality sondes and other compatible devices.
  • Standard 512kB memory gives you great flexibility for logging environmental data.
  • Sample delivery at the EPA-recommended velocity of 2 ft/sec., even at head heights of 26 feet.
  • Pump revolution counter and patented liquid detection sensor ensure accurate sample volumes - and tells you when tubing should be replaced.

Manufactured by: Teledyne ISCO
Model: ISCO 6712 Portable Waste Water Sampler

Services offered for this product

  • Sales
  • Rental
  • Installation
  • Calibration
  • Maintenance

Applications

Wastewater effluent
Stormwater monitoring
CSO monitoring
Permit compliance
Pretreatment compliance

Specification

6712 Full-Size Portable Sampler
Size (Diam x Ht) 20 x 27.0 inches (68.6 x 50.7 cm)
Weight Dry, Less Battery - 32 lbs (15 kg)
Material Plastic with UV-resistant gel coat
Power Requirements 12 VDC
Pump
Intake Purge Adjustable air purge before and after each sample.
Tubing Life Indicator Provides a warning to change pump tubing.
Intake Suction Tubing Vinyl or PTFE/FEP 3/8 inch (1 cm) ID; 3 to 99 feet (1 to 30 m) Length
Pump Tubing Life Typically 1,000,000 pump counts
Maximum Lift 28 feet (8.5 m)
Repeatability Typical ±5 ml or ±5% of the average volume in a set
Typical Line Velocity @ Head Height 3.0 ft./s @ 3 ft; 2.9 ft./s @ 10 ft; 2.7 ft./s @ 15 ft. (0.91 m/s @ 0.9 m; 0.87 m/s @ 3.1 m; 0.83 m/s @ 4.6 m)
Liquid Presence Detector Non-wetted, non-conductive sensor detects when liquid sample reaches the pump to automatically compensate for changes in head heights.
Controller
Weight 13 lbs. (5.9 kg)
Size (H x W x D) 10.3 x 12.5 x 10 inches (26 x 31.7 x 25.4 cm)
Operational Temperature 32° to 120°F (0° to 49°C)
Enclosure Rating NEMA 4X, 6 IP67
Program Memory Non-volatile ROM
Flow Meter Signal Input 5 to 15 volt DC pulse or 25 millisecond isolated contact closure.
Number of Composite Samples Programmable from 1 to 999 samples.
Clock Accuracy 1 minute per month, typical, for real time clock
Software
Sample Frequency 1 minute to 99 hours 59 minutes, in 1 minute increments. Non-uniform times in minutes or clock times 1 to 9,999 flow pulses
Sampling Modes Uniform time, non-uniform time, flow, event. (Flow mode is controlled by external flow meter pulses.)
Programmable Sample Volumes 10 to 9,990 ml in 1 ml increments
Sample Retries If no sample is detected, up to 3 attempts; user selectable
Rinse Cycles Automatic rinsing of suction line up to 3 rinses for each sample collection
Program Storage 5 sampling programs
Sampling Stop/Resume Up to 24 real time/date sample stop/resume commands
Controller Diagnostics Tests for RAM, ROM, pump, display, and distributor