Orion 3-Star pH DO Meter

It displays DO concentration or percent DO saturation and also offers a full complement of GLP related features including a 21 CFR Part 11 software option. SMART STABILITYTM and SMART AVERAGINGTM functions and a quick start guide located on the back of the unit make operation in the field even easier.

The unit offers automatic barometric pressure compensation, salinity correction and enhanced stability indicators. It can store up to 750 data logs and has the ability to turn off the data log.

Additional features include:

  • Comfortable handling
  • Backlit LCD display
  • Up to 10 methods can be stored
  • Password protection on methods
  • Optional USB outputs
  • Uses polarographic DO probe
  • IP67 waterproof rating (may be immersed for one hour with no water incursion)
Orion 4-Star pH / DO Meter Specifications
pH :
Readout Range / Resolution : -2.000 to 19.999 or 0 to 14 in aqueous solutions / 0.1,0.01,0.001
Relative Accuracy : ? 0.002
SMART STABILITY? and SMART AVERAGING? : Resolution dependent / automatic
AUTO-CAL? and/or Manual Calibration : 1 to 5 points; US/NIST, DIN, and custom buffers recognized
mV :
Range ? mV / RmV : ? 1999.9
ORP : EH autocalibration to standard hydrogen electrode
Resolution : 0.1
Relative Accuracy : ? 0.2 mV or 0.05%, whichever is greater
DO :
Range / Resolution Concentration : 0.00 to 90.0 mg/L / 0.0/0.00
Range / Resolution % Saturation : 0.0 to 600% / 0/0.0
Relative Accuracy Concentration / % Saturation : ? 0.5% ?1 digit
SMART STABILITY? and SMART AVERAGING? : Resolution dependent / automatic
Auto Calibration or Manual Input : Water Saturated Air and Air Saturated Water, Winkler
Calibration to Winkler Titration for BOD : Available as needed
Salinity Factor : 0 to 45 ppt
Barometric Pressure Range : 450 to 850 mm Hg
Probe Type : Polarographic
TEMPERATURE :
Range / Resolution : -5 to 105 ?C / 0.1 up to 99.9 ?C, 1.0 over 99.9 ?C
Relative Accuracy : ? 0.1 ?C
Temperature Compensation : DO (auto); pH (auto/manual)
Display : Custom backlit LCD, for easy viewing
INPUTS / OUTPUTS :
Sensor Inputs : DO-miniDIN - Locking; ATC-miniDIN-Locking; pH/ISE - BNC; (all waterproof)
Communications : Waterproof bi-directional RS232 - Stereo jack, optional USB interface Cable
SPECIAL GLP SOFTWARE FEATURES :
# Of Data Logging Points : 750 points by time, stability or manual read with time/date stamp
Calibration Log : Last 10 Calibrations
User Method Storage : 10 Saved
Advanced Self-test and Diagnostics : YES
GLP Electrode Verification : YES
Password Protection : YES
Internet Software Updates / Downloadable Methods : YES
21 CFR Part 11 Software Package : YES
Power : 4 x AA batteries, powerless date protection
Regulatory and Safety : CE, CSA, UL, T?V, FCC Class A limits
ENVIRONMENTAL OPERATING CONDITIONS :
Ambient Operating Temperature : 5 to 45 ?C
Relative Humidity : 5 to 85% Non-Condensing
IP Rating : IP67, waterproof and dustproof
Warranty : 36 months (from date of purchase)
* Above data generated under optimal/controlled circumstances
Download Product Brochure Download PDF

Select Options